A New Standard in Real-Time
Aerosol Monitoring
The E-SAMPLER is a comprehensive system
that provides both real-time monitoring and filter-based sampling capability. Whatever
your monitoring objectives, the E-SAMPLER will provide accurate, dependable and relevant
data.
Features
Aluminum Weatherproof Enclosure
Programmable Auto Zero & Span
Integral 47mm Filter
Adjustable Sample Flow
Actual Flow Control
Programmable Inlet Heater
Sheath-Air Protected Optics
Auto-Ranging
10-Channel Data Logger
Meteorological Sensor Inputs
Internal Battery |
|
 |
Dual
Technology
The E-SAMPLER combines real-time
measurement utilizing light-scatter technology, and the ability to conduct gravimetric and
chemical analysis.
Particulate loading on the filter does not reduce the performance of
the light-scattering detector. Flow, temperature and pressure parameters are continuously
measured, allowing the on-board microprocessor to control the actual sampling rate. |
|

|
Principle
The E-SAMPLER provides real-time
measurement through its near-forward light scattering technique. The on-board sample pump
draws air at 1.0 to 3.5 lpm into the sensing chamber. A state-of-the-art detector measures
the intensity of scattered light, which is proportional to the particulate concentration. |
Maintenance
In addition to the 47mm sample filter,
the E-SAMPLER has two internal filters. One protects optical surfaces and the other
prevents damage to the flow controller. Both filters are easily accessible from the front
panel. Filter life will exceed one year of operation in the harshest of conditions. |
|

|
A curtain of clean air continuously protects all optical
surfaces within the detector. This sheath-air allows the instrument to be used in adverse
environments without performance degradation.
Operation
The E-SAMPLE is
rugged, portable and easy to use. Initial set-up is simple. Once configured, simply
turning the monitor on will begin operation. The unit will continue to operate until user
intervention, or the battery has failed.
Auto-Zero and
Auto-Span ensure that the data collected will be of the highest quality. Both Zero and
Span can be operated manually, or programmed to occur at user specified intervals.
The E-SAMPLER
can also be programmed to operate at defined intervals (start / stop times). Other
parameters that may be controlled by the user include recording periods and averaging
time.
Specifications
| Ranges: |
0.5, 1, 10, 65 mg/m3 |
| Sensitivity: |
0.001 mg/m3 |
| Sampling Rate: |
1 second |
| Flow Rate: |
1 - 3.5 lpm |
| Accuracy: |
8% of NIOSH 0600 |
| Precision: |
2% of reading |
| Auto-Cal Interval: |
0.25 - 24 hours |
| Averaging Period: |
0.1 - 60 minutes |
| Display: |
4 x 20 LCD |
| Outputs: |
1, 2.5 or 5 VDC |
| Serial Port: |
RS232 |
| Data Capacity: |
12,000 records |
| Alarm: |
Contact Closure |
| Inlet Heater: |
RH Controlled |
| Power: |
12VDC |
| Consumption: |
275mA |
| Battery Type: |
Lead Acid |
| Battery Life: |
40 hours |
| Operating Range: |
-10º to +50º C |
| Dimensions: |
267 x 235 x 145 mm |
|