LAND COMBUSTION

Model 4500 Mark II
EPA Complaint Opacity Monitor

 

The versatile Model 4500 Mark II is a dual-pass transmissometer that offers a reliable and effective solution to the problem of dust measurement in industrial flue gases.

With no moving parts, the optical measuring system of the Model 4500 is extremely reliable. The patented 'Flood LED' technique enhances the stability of this instrument by compensating for any drift in the photo-electric detectors.

The Model 4500 Mark II wholly complies with the latest EPA performance specifications (CFR 40 Part 60, App. B, PS-1).

Features:

An electronically modulated (LED) light source
No moving parts during measurement cycle
Automatic zero and span calibrations
A patented flood technique for continuous photo-detector compensation
Zero sensitivity to changes in ambient light
Automatic measurement and correction of optical contamination
Simultaneous opacity and dust monitoring capability
A built-in calibration audit filter fixture (zero-jig)
Independently configurable analog outputs
Instrument status and exceeding limit relays
RS232 / RS422 / MODBUS interface
Continuous and comprehensive self diagnostics

Specifications:

Transceiver & Retro-Reflector
Technique Dual Pass Transmissometer
Wavelength 565 ± 20 nm
Light Source High Intensity LED (lifetime > 5 years)
Accuracy ± 2% Opacity
Repeatability ± 0.5% Opacity
Resolution 0.1% Opacity
Drift < 1% per month
Angle of Projection < 2 degrees
Angle of View < 2 degrees
Response Time < 1 second to 90% of final value
Operating Temperature -20º to +55º C Standard (standard)
Max.Flue Gas Temperature +600º C (standard)
Measurement Pathlength 0.7m - 10.0m (standard)
Local Control Unit (LCU)
Output Range 0 - 5% to 100% (selectable in 1% increments)
Display 2 x 16 character LCD
User Interface Tactile membrane keypad with password protection
Available Measurements Opacity, transmission, optical density (extinction) or dust (as mg/m3)
Outputs 4 - 20mA (individually isolated)
Averaging Selectable from 1 to 59 minutes or 1 to 24 hours
Calibration On-board zero and upscale calibration function. Automatically initiated by software (selectable interval 0 to 48 hours) and/or via remote trigger input
Calibration Alarms Indication of excessive zero drift, span drift, and optical contamination
Relays 6 isolated changeover relays (240VAC - 1A or 24VDC - 5A) for: high opacity 1, high opacity 2, malfunction, maintenance, zero calibration in progress, span calibration in progress
Indicators LEDs for power on, system OK, calibration in progress, calibration error, maintenance in progress, opacity limits, optical contamination threshold, air purge failure
Accessories Single or dual blower air purge units, weatherhoods, NIST traceable calibration filters.

 